Features and Benefits
The DTS-100D and DTS-100DF are CE Marked liquid dielectric test sets, typically used for testing the breakdown voltage level of insulating oils, hydraulic oils, and other dielectric fluids per the most common world standards including ASTM D-877, ASTM D1816, and IEC 156. Three user-selectable automatic voltage rates-of-rise, and automatic voltage shutdown with the meter indicating the breakdown voltage of the insulating liquid under test. Its durable aluminium construction, small footprint, a removable oil drip tray and rugged design make our DTS series ideal for field, factory, or laboratory use.
Three motorised rates of rise: 500V/2000V/3000V per second
Arc detection with automatic shutdown and failure indicator
Window panel for test observation
Accessory outlet located within test chamber
One-piece portable design
Zero Start safety and test chamber interlock provision
Specifications
Input:
120 Vac, 60hz, 7A (DTS-100D)
230 Vac, 50/60hz, 3A (DTS-100DF)
Output:
0-100 kVac, 800 VA resistive load
Meter Accuracy:
+/- 2% of full scale
Oil Test Cells:
TCD-3 for ASTM D877, TCD-5 for ASTM D1816, TCD-12 for IEC 156
Optional Test Cells (Sold Seperately):
TCD-N3 for ASTM D877, TCD-N5
for ASTM D1816, TCD-NI2 for IEC 156
Operating Temperature:
14° to 104° F, -10° to 40° C
Size & Weight:
(W x D x H)
18.5 x 16.5 x 16.5 in., 100lb
470 x 419 x 419 mm, 45kG
Output Termination:
Dual high voltage sections with brass corona spheres in test chamber
Test Cells
TCD-N3 – For ASTM D877, flat disc electrodes, 625mL sample size, 0.100-inch gap adjustment gauge
TCD-N5 – For ASTM D1816, spherical electrodes, 625mL sample size, 0.100, 0.080, and 0.040-inch gap adjustment gauges, includes motorised stirring assembly
TCD-NI2 – For IEC 156, spherical electrodes, 625mL sample size, 0.100, 0.080, and 0.040-inch gap adjustment gauges
Optional Accessories
PN 32-0391 Reusable Shipping Container