PFT Series - PFT-301CM(F)

PFT Series - PFT-301CM(F)

PFT-301CMF.png

Features and Benefits

0 – 30 kV AC Power Frequency AC Hipot

The PFT-301CM and PFT-301CMF are from our popular PFT Series of field portable AC hipot test sets. These hipots provide continuously adjustable AC voltage used for a pass/fail, AC over-voltage withstand test on high voltage apparatus, like switchgear, circuit breakers, reclosers, vacuum bottles, hot sticks, bus duct, and any other apparatus needing an AC test voltage. The PFT Series offers the most field portable, affordable, and easiest to operate AC hipots at this voltage and power. Equipped with a two-range voltmeter, three-range current meter and guard circuit for precise leakage current measurements. For bench-top requirements see the ABT Series. Testing loads that require higher power, see our ALT, FPA, HPA and PAR lines of AC dielectric test sets.

  • Convenient one-piece field portable design

  • Secondary connected single range Voltmeter: 0 – 30 kV AC

  • Single Range Current Meter: 0 – 40 mA AC

  • Fixed overload set to 120% of variable transformer rated output current

  • Transit protected meters prevent meter damage between test sites

  • 20 ft shielded output and return cables

  • Zero Start and External safety Interlock provisions

Specifications

Input:

(PFT-301CM) 120 V, 60 Hz, 10 A
(PFT-301CMF) 230 V, 50/60 Hz, 5 A

Output:

0-30 kV AC, 1 kVA resistive load
1 kVA capacitive load, 33 mA current

Duty:

1 kVA: 1 hour on, 1 hour off
700 VA: Continuous

Voltmeter:

3.5”, scaled 0-30 kV AC, ±2% F.S.

Current Meter:

3.5”, scaled 0-40 mA, ±2% F.S.

Meter Accuracy:

2%

Distortion:

Less than 5%

Size & Weight:
(W x D x H)

17 x 11.5 x 14 in., 45 lb.
431 x 292 x 365 mm, 20 kg

Output Termination:

8 ft. (2.5 m) shielded EPR output cable with alligator clamp

Scope Of Supply:

20ft black ground lead, external interlock plug, operations
manual, calibration certificate

Optional Accessories

Hand & Foot Safety Switches

  • PN 90-410

  • PN 90-405

Safety Ground Sticks

  • PN 88-150

  • PN 88-113

Optional Upgrades

  • Digital Metering – Add Suffix “D”