Features and Benefits

HVI’s smallest VLF model, the VLF-30CM(F) approaches the size and cost of a DC hipot and is designed for quick and easy short-run cable testing. It can test up to 0.4 mF of load, or up to 4000 feet (1200 m) of a typical 15 kV class cable.

Specifications

Input:

120 V, 50/60 Hz, 5 A peak, 2.5 A average
(F) 230 V, 50/60 Hz, 3 A peak, 2 A average

Output:

0 – 30 kV AC peak, 0.1 Hz, sinusoidal

Duty:

Continuous

Load Rating:

0.4 µF @ .1 Hz

Voltmeter:

3.5″, -30 – 0 – +30 kV AC, 2% FS accuracy

Current Meter:

3.5″, 0 – 50 mA peak, 2% FS accuracy

Output Termination:

Shielded output cable – 20 ft. (6 m)

Size & Weight:
(W x D x H)

14.25 x 11.25 x 21.75 in., 85 lb.
362 x 286 x 552 mm, 39 kg